·
Is there sample carryover?
·
Are there sample interferences or complexing
agents present?
·
Is the reference electrode junction contaminated?
SLOW RESPONSE (
·
Was the electrode stored in the wrong solution?
·
Was the electrode poisoned by the sample?
OUT OF RANGE
·
Is the electrode plugged into the controller properly?
·
Is there a reference electrode present?
·
Is there enough fill solution left in the reference electrode?
·
Is the electrode placed in the sample solution?
LOW SLOPE OR NO SLOPE
·
Are the standards too old, contaminated, or made wrong?
·
Has the sample pH been adjusted properly to the operating range of
the electrode?
·
Is there an air bubble on electrode surface?
·
Is the controller okay?
·
Is there enough fill solution left in the reference electrode?
NOISY RESPONSE (
·
Is the controller grounded?
·
Is there an air bubble on electrode surface?
·
Is the controller operating properly?
·
Is there enough fill solution left in the reference electrode?
DRIFTY RESPONSE (
·
Is there excessive leaking at the reference electrode junction?
·
Is the reference electrode junction clogged?
·
Was the sensing membrane poisoned by the sample?
·
Are there temperature problems?
·
Is the sample too concentrated?
·
Does the sensing membrane need conditioning?
INACCURATE (BUT CALIBRATION IS OK)
·
Are the standards incorrect?
·
Was the sample pH adjusted properly?
·
Is there sample carryover?
·
Are there sample interferences or complexing
agents present?